Data Logger Specifications

Alarm Type Audible/Visual (Phone, Text, and Email Alarms via DicksonOne Subscription)
Ambient Operating RH Conditions 0 to 95% RH non-condensing (Logger Only)
Ambient Operating Temperature Conditions TWE: 32.0 to 158.0F (0.0 to 70.0C)
TWP: 32.0 to 140.0F (0.0 to 60.0C)
Antenna Type Internal
Backup Battery 70 Hours
Battery Type Rechargeable Backup Battery
Cable Details Up to 5 Probe Extensions (A898) can be used (50ft Total)
Calibration Options N100, N300, N400, N995
Data Capacity Approx 1,000,000 sample points (backup)
Dimensions 8.5 x 1.75 x 7.0" (21.0 x 4.5 x 18.0cm)
Display Dimensions 8" Diagonal
Display Type LCD Touchscreen
Display Update Rate Sample Interval Dependent (User Selectable)
Download Type WiFi/Ethernet (DicksonOne Cloud), USB (must be formatted at FAT32)
Enclosure Case Front: Polycarbonate; Case Back and Pods; ABS
Included Accessories AC Adapter, Mounting Bracket and Kickstand
POE: Input Current Peak TWP: (At 48V) 235mA
POE: Input Voltage Range TWP: 36.0 to 57.0V
Mounting Options Mounting Bracket (Included with Touchscreen)
PC Requirements Windows 7 or more recent
Power Consumption TWP: Up to 9W
Power Cord 6 ft
Power Source 100-240 VAC 50/60Hz 12V=1A
Required Software DicksonOne SaaS Package, DicksonWare VERSION TBD
Resolution LCD TBD
Sample Interval User Selectable Sample Intervals: 60, 90 seconds / 2, 5, 10, 15, 30, 60 minutes. (10 and 30 seconds optional, please call for details)
Sensor Type Replaceable Sensor (See Accessories for more information)
Unit Weight 1.45lbs
Units/Pkg 1
Warranty 12-month
IP Rating 21





Sensor Specifications

Sensor Description SKU Specifications
Ambient Temperature & Humidity Replaceable Sensor RTRH
RTRH-R
Ambient Operating RH Conditions : 5 to 95% RH (non-condensing)
Humidity Accuracy: +/-2.0% RH from 5.0 to 95.0% RH
Humidity Range: 5 to 95% RH (non-condensing)
Temperature Accuracy: +/-0.8F from 20.0 to 120.0F 1.8F (+/-0.4C from -6.7 to 50.0C 1.0C)
Temperature Range:-40.0 to 185.0F (-40.0 to 85.0C)
Single Thermistor Replacaebale Sensor (in buffer) RTHM Temperature Accuracy: +/-0.9F from -58.0 to 68.0F 0.0F (+/-0.5C from -58.0 to 20.0C 0.0C)
Temperature Accuracy (Remainder of Range): +/- 1.2F (+/-0.9C) from 69 to 122F (21 to 50C); +/-1.8F (+/-1C) from 123-158F (51- to 70C)
Temperature Range: -58.0 to 158.0F (-50.0 to 70.0C)
Dual Thermistor Replacaebale Sensor (in buffer) RTHM2 Temperature Accuracy: +/-0.9F from -58.0 to 68.0F (+/-0.5C from -50.0 to 20.0C)
Temperature Accuracy (Remainder of Range): +/- 1.2F (+/-0.9C) from 69 to 122F (21 to 50C); +/-1.8F (+/-1C) from 123-158F (51- to 70C)
Temperature Range: -58.0 to 158.0F (-50.0 to 70.0C)
K-Thermocouple Replaceable Sensor RKTC Temperature Accuracy: +/-1.8F from -300.0 to 122.0F (+/-1.0C from -185.0 to 50.0C)
Temperature Accuracy (Remainder of Range): +/-3.8 (+/-2.1C)
Temperature Range:-300.0 to 2000.0F (-185.0 to 1093.0C)
Dual K-Thermocouple Replaceable Sensor RKTC2 Temperature Accuracy: +/-1.8F from -300.0 to 122.0F 3.8F (+/-1.0C from -185.0 to 50.0C 2.1C)
Temperature Range: -300.0 to 2000.0F (-185.0 to 1093.0C)
Platinum RTD Replaceable Sensor RRTD Cable Length: 6.0ft (182.0cm)
Temperature Accuracy: +/-0.5F from -148.0 to 350.0F (+/-0.3C from -100.0 to 176.0C)
Temperature Range:-148.0 to 350.0F (-100.0 to 176.0C)
Ambient Temperature Replaceable Sensor RTMP
RTMP-R
Temperature Accuracy: +/-1.2F from 20.0 to 122.0F (+/-0.7C from -7.0 to 50.0C)
Temperature Range: -22.0 to 122.0F (-30.0 to 50.0C)
Differential Pressure Replaceable Sensor (2") RS080 Differential Pressure Accuracy: +/-3.0% from -2.0 to 2.0inH20
Resolution: 0.000
Differential Pressure Replaceable Sensor (0.5") RS081 Differential Pressure Accuracy: +/-3.0% from -0.5 to 0.5inH20
Resolution: 0.000