Palmer Aluminum Case Pressure and Temperature Chart Recorders provide proven reliability for accurate measurement and recording of pressure and temperature in many applications. Since 1962 we have manufactured and supplied recorders for industry in the U.S. and around the world. Palmer Recorders are designed for indoor or outdoor use, and provide long service life, even in harsh conditions. Many of our Recorders are still in service after 50 years!
Temperature and Humidity Datalogger SK-LTHIIa series
:
Sato Datalogger SK-L200THIIa for temperature and humidity measurements.
CD-ROM for data analysis software and real-time transmission software is attached as accessories.
CD-ROM for data analysis software and real-time transmission software is attached as accessories.
Logged data can be easily displayed and analized on a PC using the software.
Three types of probes are sold separately.
Three types of probes are sold separately.
Features
|
Standard and Optional Probes | |||||||
|
|||||||
Model | SK-LTHIIα-1 Plug-in type |
SK-LTHIIα-2 w/ sensor cord |
SK-LTHIIα-3 Stick type w/ sensor cord |
||||
Cat No. | No.8176-00 | No.8177-00 | No.8178-00 | ||||
Sensor (Temp.) | Thermistor | ||||||
Sensor (Hum.) | High Polymeric Resistance change type | ||||||
Measuring range (Temp.) | -10 to 60.0°C | ||||||
Measuring range (Hum.) | 20.0 to 98 % rh (at 23 °C) | ||||||
Accuracy (Temp.) |
±0.5°C at 20.0 to 30.0°C ±1°C at other than above |
±0.5°C at 15.0 to 35.0°C ±1°C at other than above |
|||||
Accuracy (Hum.) |
±3% rh at 20.0 to 30.0°C (30 to 90% rh) ±7% rh at less than 10.0°C, higher than 50.0°C ±5% rh at other than above |
±3% rh at 15.0 to 35.0°C (30 to 90% rh) ±7% rh at less than 10.0°C, higher than 50.0°C ±5% rh at other than above |
|||||
Dimensions | (W) 25 x (H) 70 x (D) 20 mm | (W) 18 x (H) 51 x (D) 10 mm | Ø8 x (L)35 mm | ||||
Cord length | --- | 1.5 m | 1.5 m | ||||
Weight | about 15 g | about 65 g | about 69 g | ||||
Materials | ABS resin | ABS resin, PVC resin | ABS resin, PVC resin, Polyacetal resin | ||||
Accessories | Test Result |
Main units | |
Model | SK-L200THII α (temperature and humidity measurement type) |
Display range | -15 to 65.0°C (depending on a probe connected) |
10 to 99.9% rh | |
Resolution(Temp.) | 0.1°C at -9.9 to 65.0°C 1°C at -10°C or less |
Resolution(Hum.) | 0.1% rh at 15.0 to 99.9% rh 1%rh at less than 15.0% |
Accuracy (Temp) | ±(0.1°C + 1 digit) |
Accuracy (Hum.) | ±(0.1% rh + 1 digit) |
Probes connected | SK-LTHII α series probes |
Sampling (display) | about 1 sec. |
Display functions | TIME (display time), INT (log interval) REC (during logging), RES (during preset standby), OUT (during real-time transmission), LCD bars (used memory) CALL (call for the current measurement data during logging) Current temperature, Current humidity, Current date and time, low battery symbol, Temp. upper limit (Hi), Temp. lower limit (Lo), Hum. upper limit (Hi), Hum. lower limit (Lo), Temp. alarm ON (Alarm), Hum. alarm ON (Alarm), logging option, setting value of upper and lower limits, error message, ID number |
Memory capacity | 8100 lines of data, but depending on the logging option (four logging potions: Once, Repeatability, Ends at, By page) |
Log interval | 1 sec., 2 sec., 5 sec., 10 sec., 15 sec., 30 sec. 1 min., 2 min.,5 mi.,10 min., 15 min., 30 min., 60 min., 90 min. |
Communication | USB |
Operation ambient | -10 to 60°C, no condensing |
Storage ambient | 0 to 50°C |
Power requirement | "AAA" size Alkaline battery x 2 |
Battery life | . Under continuous measurement mode : more than 500 hours . Under logging mode: possible to log 8100 data with every log interval (ex.: about 506 days with log interval of 90 minutes.) |
Dimensions | (W) 100 x (H) 73 x (D) 22 mm |
Weighs | about 123 g (exclusive of batteries) |
Materials | ABS resin |
Accessories | CD-Rom (Data analysis and realtime software), USB cable (1.5 m) , USB port cap (factory set), hand strap, "AAA" battery x 2 |