Sato SK-8950 is a waterproof type Infrared thermometer having the circle laser marker that surrounds the measuring area with red laser. The Waterproof/dustproof property is IP 54. The measuring range is -60 to 550°C. Emissivity is settable with 0.01 steps between 0.10 to 1.00
Fiber Optic Thermometer FL Series
:
Anritsu FL Series can measure temperatures in environments subject to high-frequency waves, microwaves, or high voltages-areas where electric thermometers such as thermocouples, platinum resistance temperature detectors, and thermistors are typically ineffective.
・Capable of measurements in environments subject to high-frequency waves or high voltages.
・A fiber optic thermometer offering a wide measurement range.
・Analog output capability allows use for machine temperature control.
・Measurement data can be imported into a PC using software* bundles with the thermometer.
* for Microsoft® Windows® 7, Windows® 8.1, Windows® 10 (for English)
FL-2000 (1ch model)(Temperature probes not included) |
|
FL-2400 (4ch model)(Temperature probes not included) |
FS100-*M (General-purpose type)
Applications involving microwaves, high magnetic fields, certain food preparation processes, power lines, and electric circuit components
temp. range : -190°C to 260°C | repairable |
FS150-*M(High durability general-purpose type)
The durable type which sensing part of FS100 made of the same thickness as the cable.
temp. range : -190°C to 260°C | repairable |
FS100H-**-*M(Liquid and chemical-resistant type)
Oils or chemicals.
temp. range : -190°C to 260°C | repairable |
FS150H-**-*M(High durability liquid and chemical-resistant type)
The durable type which sensing part of FS100H made of the same thickness as the cable.
temp. range : -190°C to 260°C | repairable |
FS200-*M(Chemical-resistant type)
Chemicals.
temp. range : -30°C to 200°C | repairable |
FS300-*M(Light-resistant type)
Measurements performed in laser beam, plasma, or other intense light
temp. range : -190°C to 260°C | repairable |
FS400-*M(High temperature type)
Various materials, excluding liquids, in semiconductor
temp. range : -20°C to 400°C | repairable |
FS500-*M(Needle type)
Semisolid substances, e.g., food
temp. range : -60°C to 260°C | repairable |
FS600-*M(Plastic type)
Animal tests using hyperthermia equipment or MRI or NMR systems
temp. range : 0°C to 80°C | Unrepairable |
Optional accessories
FK1-*M(Extension cable)
General-Purpose extension cable for normal use.
repairable |
FK2-*M(Extension cable)
Heavy-duty extension cable for rough environments.
repairable |
FA1 (Optional adapter)
Con connect a probe to an extension cable.
Unrepairable |
FN1-T(sealant material : TFE) (Compression fitting)
Compression fitting used to insert probe into a vacuum device or pressure vessel.
NPG Size | NPT1/8 |
Pressure rating | 6.6×10-4Pa to 11.0MPa at 20°C using stainless steel rod |
FN1-T(sealant material : TFE) | Unrepairable |
FN1-V(sealant material : Viton) (Compression fitting)
Compression fitting used to insert probe into a vacuum device or pressure vessel.
NPG Size | NPT1/8 |
Pressure rating | 6.6×10-4Pa to 11.0MPa at 20°C using stainless steel rod |
FN1-V (sealant material : Viton) | Unrepairable |
Model | FL-2000 | FL-2400 | |
Channel | 1 | 4 | |
Measurement Range | -195.0 to 450.0°C | ||
Resolution | 0.1°C | ||
Accuracy | Calibration not executed | -195.0 to 399.9°C:±2°C 400.0 to 450.0°C:±5°C | |
Calibration executed |
±0.5°C (within ±50°C of caribration point) |
||
Temperature coefficient | ±(0.01% of F.S./°C) Outside 25℃±5℃ | ||
Operating environment |
Temperature : 0 ~ 40°C Humidity : 15 ~ 80%RH(non-condensing) |
||
Storage environment |
Temperature : -10 ~ 50°C Humidity : 10 ~ 85%RH(non-condensing) |
||
Display | Height of character : 15mm |
Height of character : 8mm RED |
|
Display update | 1sec | ||
Light source | Blue LED | ||
Power supply |
Please specify a (choose from six AA-size alkaline batteries |
Please specify a 100VAC, 115VAV, or 230VAC) |
|
Power consumption | approx. 12 VA | approx. 40 VA | |
Memory capacity |
UP to 15,000 data, one second interval (synchronized with display) |
No memory | |
Interface |
RS-232C Transfer rate : 9600bps Data formation : 7bit, odd parity, 1 stop bit |
||
Calibration |
One-point calibration (using real temperature within a measured area) |
||
Analog output | Output rate | 10mV/°C | |
Resolution | 1mV | ||
Accuracy | ±(0.2% of F.S) at an ambient temperature of 25°C±5°C | ||
Temperature coefficient | ±(0.01% of F.S./°C) | ||
Dimensions |
210(W)×88(H)×230(D) mm Protruded parts excluded |
320(W)×99(H)×280(D) mm Protruded parts excluded |
|
Weight | approx. 3kg | approx. 5kg | |
Supplied accessories |
Power cable(two-wire plug attached)×1 Fuse(2A mini fuse)×1 Analog output cable×1 communication cable×1 PC software AMS-850 (CD ROM)×1 User's manual(main unit and software)×1 |
Power cable(two-wire plug attached)×1 Fuse(2A mini fuse)×1 Analog output cable×4 Communication cable×1 Conversion cable (RS-232C→USB)×1 PC software AMS-850 (CD ROM)×1 User's manual(main unit and software)×1 |
Operational environment of software (AMS-850)
OS | Microsoft® Windows® 7, Windows® 8.1, Windows® 10 (for English) |
Memory | 64MB or more (Operational environment of corresponding OS) |
HDD | Min.5MB of free space (Operational environment of corresponding OS) |
Monitor | Min.resolution of 800×600 pixels. |
* All PCs under the recommended environment are not guaranteed to work successfully with AMS-850. * AMS-850 can be used only by the user that has a system administrative right (Administrator) * AMS-850 does not support Mac OS. * Microsoft® Windows® 7, Windows® 8.1 and Windows® 10 are registered trademarks of Microsoft Corporation in the United States and other countries. |