Anritsu Thermometer HD series handheld thermometers allow easy temperature measurement.
These general-purpose thermometers can be used for on-site temperature measurement and various other applications.
Waterproof Infrared Thermometer
:
Sato SK-8950 is a waterproof type Infrared thermometer having the circle laser marker that surrounds the measuring area with red laser. The Waterproof/dustproof property is IP 54. The measuring range is -60 to 550°C. Emissivity is settable with 0.01 steps between 0.10 to 1.00
Features
|
Specifications
Cat. No. | No.8269-00 |
Model | SK-8950 Waterproof Type Infrared Thermometer with circle laser marker |
Measuring range | -60°C to 550°C |
Measurement accuracy | ± 2°C or 2% reading at 0.0 to 550°C ± (2°C + 5% reading) at other than above range Conditions: Operation ambient at 23°C±3°C Emissivity: 0.95 |
Resolution | 0.1°C |
Emissivity | Settable with 0.01 step in the range ε= 0.10 to 1.00 (Refer to the emissivity table) |
Distance to Spot size | D:S = approx. 12:1 (D: distance, S: diameter of measurement area) |
Sensing element | Thermopile |
Spectral response | 8 to 14μm |
Laser marker | Red laser circle (class II laser product) Wavelength: 650nm Output: less than 1mW |
Ingress protection code | JIS C 0920 IP54 |
Operating ambient | 0 to 50°C, less than 85%rh (no condensing) |
Storage ambient | -10 to 50°C (no condensing) |
Battery life | At continuous measurement: approx. 14 hours * Conditions: use of alkali batteries, laser marker and backlight are ON |
Dimensions. Weight | approx. (W)42 x (H)143 x (D)80mm approx. 196g (including batteries) |
Accessories | 2 pcs. of AAA alkali battery (LR03), Instruction Manual, one neck strap |
In SK-8950, the distance and the diameter of measurement area (D:S) is approx. 12:1 The distance and spot size (The measuring spot is defined on the basis of the area that receives 90% or more of the energy) |
Relation of measurement area and the position of circle marker |
* Cautions Standard for Safe Use of Laser Products (JIS C6802): "Class II" is defined as follows: A visible light (of wavelengths in the visible region: 400 nm through 700 nm) with an output level (approx. 1 mW or less) that is normally regarded harmless with respect to humans' physical defense capabilities. |