EDL Practi-Cal (PRI) Dry Well / Surface Calibrator (PRI) allows for easy testing and verification of virtually any size or shaped surface and insertion-style sensor – in one compact package. Utilizing an interchangeable dry well system, this calibrator has a uniform flat surface alongside a custom drilled insertion dry well. The PRI is ideal for use in the lab or in the field. It weighs in at under 30lbs., and can be taken virtually anywhere.
Surface Temperature Calibrator
:
Anritsu ACSⅡ series products are temperature calibration systems for surface temperature and internal temperature measuring devices.
Temperatures can be set across a wide range, from room temperature to 500°C. These calibration systems are designed for calibration or to check temperature measuring probes.
ACSⅡ-1000![]() For Surface Temperature Max. 300°CSurface Range: Rt to 300°C |
ACSⅡ-2000![]() For Surface Temperature Max. 500°CSurface Range: Rt to 500°C |
ACSⅡ-3000![]() For Internal Temperature Max. 500°CInternal Range: Rt to 500°C |
ACSII-4000![]() For Suface and Internal Temperature Max. 500°CSurface Range: Rt to 500°C |
Overall Accuracy
| Model | ACSII-1000 | ACSII-2000 | ACSII-3000 | ACSII-4000 | ||
| Calibration Object | Surface | Surface | Internal | Surface | Internal | |
| Range | Rt to 100°C | +0.5°C -0.7°C |
+0.5°C -0.7°C |
±0.26°C | +0.4°C -0.6°C |
±0.26°C |
| 100.1 to 200.0°C | +0.5°C -1.1°C |
+0.5°C -1.1°C |
±0.32°C | +0.3°C -0.9°C |
±0.32°C | |
| 200.1 to 300.0°C | +0.5°C -1.5°C |
+0.5°C -1.5°C |
±0.37°C | +0.1°C -1.1°C |
±0.37°C | |
| 300.1 to 400.0°C | ― | +0.4°C -2.0°C |
±0.43°C | -0.2°C -1.5°C |
±0.43°C | |
| 400.1 to 500.0°C | ― | -0.1°C -2.9°C |
±0.48°C | -0.9°C -2.2°C |
±0.48°C | |
Control Unit
| Model | ASCII-1000 | ACSII-2000 | ACSII-3000 | ACSII-4000 | |
| Range | Rt to 300°C | Rt to 500°C | Rt to 500°C | ||
| Control Principle | PID auto-Tuning Control | ||||
| Control | Input | Pt100Ω φ3.2 | |||
| Accuracy | ±(0.2% of reading)℃ | ||||
| Display | 8mm high LED display | ||||
| Reference Sensor |
Input | - | Pt100Ω Φ4.8mm JIS C 1604 | ||
| Accuracy | - | ±(0.05% of reading+0.05)℃ | |||
| Display | - | 12.5mm high LED display | |||
| Power supply | AC100V~240V * Please select | ||||
| Power Consumption | app.160W | app.700W | app.1600W | ||
Calibration Unit
| model | ACSII-1000 | ACSII-2000 | ACSII-3000 | ACSII-4000 |
| Testing Surface (mm) | φ78mm | φ100mm | - | □150mm |
| Test Point | - | 8 holes based on specified dimension | ||
| Temp. Distribution Surface | ±0.1°C | - | ±0.1°C | |
| Temp. Distribution Internal | - | ±0.1°C | ||
| Heater | 150W | 600W | 1500W | |
| Overheating prevention mechanism | Bimetal breaker | |||
| Heating time | Rt → 150°C 25min Rt → 300°C 50min | Rt → 250°C 30min Rt → 500°C 60min | Rt → 250°C 60min Rt → 500°C 150min | |
| Cooling time | 300 → 150°C 80min 150 → 50°C 120min | 500→250°C 40min 250→50°C 90min |
500→250°C 220min 250→100°C 450min |
|
Supplied accessories
| Model | ACSⅡ-1000 | ACSⅡ-2000 | ACSⅡ-3000 | ACSⅡ-4000 |
| Supplied accessories | Power cable,Heater cable, Sensor cable, Manual,Test report | |||
| Protective cover for calibration block | Standard sensor for calibration (Pt 100Ω platinum resistor) |
|||